Method 21.1 High Pressure Liquid Chromatography (HPLC)
Purpose: To determine pungency levels in capsicums and their
oleoresins.
A. Apparatus:
- High pressure liquid chromatograph, pump flow rate (0.6 to 1.8
ml/min).
- Detector
UV at 280 nm, flow cell path (1 x 10 mm), greater
than 700 ppm capsaicin.
Fluorescence at 288 excitation, 320 emission, flow cell path
of 10 x 1 micron, less than 700 ppm capsaicin.
- Column: C18, 10 micron spherical, 4.6 mm x250 mm.
- Rheodyne syringe loading sample injector with 20 microliter loop.
- 100 ml volumetric flask.
B. Reagents:
- Acetonitrile, HPLC grade.
- Dioxane, HPLC grade.
- HPLC grade water.
- HPLC grade methanol.
- 2% perchloric acid.
- 95% ethanol, saturated with sodium acetate.
- HPLC grade sodium acetate.
- Capsaicin. (Pharmaceutical quality).
C. Procedure:
- Weigh sample into 100 ml. volumetric flask:
- 10.00 g for pungencies from 0 to 1000 ASTA Heat Units
- 1.00 g for pungencies from 1,000 to 10,000 ASTA Heat Units
- 0.10 g for pungencies greater than 10,000 ASTA Heat Units
- Bring to 100 ml. with 95% ethanol saturated with sodium acetate.
- Cap and place on hot plate at 60 C or water bath for three hours, swirling occasionally.
- Swirl and allow solids to settle.
- Fill sample loop and elute with a mix* of: 48.4% methanol, 30.2% HPLC grade water, 13.3%
dioxane, 7.9% acetonitrile, 0.2% of 2% perchloric acid.
- The first peak is the pigment peak, the second peak is capsaicin, the third peak is
capsaicinoid.
- Prepare capsaicin standards at 0.01, 0.10, 1.0 mg./ml. (label 10, 100 and 1,000 ppm
capsaicin) in 95% ethanol, saturated with sodium acetate.
*For low pungency capsicums (less than 700 ppm capsaicin):
- Prepare solvent mixture A (5.4% acetonitrile, 3.6% dioxane, 0.7% perchloric acid, 90.3%
water) and B (9.0% acetonitrile, 18.0% dioxane, 73.0% methanol).
- Program gradient separation: At 0 min. (0% solvent B); at 2.5 min. (increase B from 0 to
60% in 5 min.); 7.5 to 12.5 min. (isocratic at 60%); 12.5 min. (60% to 100% B).
Calculation:
- Total capsaicin area=Area peak 2+.82 Area peak 3.
- Capsaicin (ppm) = Total capsaicin area of sample x ppm standard x 100 / Total capsaicin
area of standard xg.
- Sample ASTA Pungency=ppm capsaicin
- Pungency (Scoville) =ppm capsaicin x 15 (assume Capsaicin standard at 15,000,000
Scoville)
ppm capsaicin x 15 = Scoville units (assumes capsaicin = 15,000,000 Scoville)
|